CHEMOTRODE BRIDGE,
CHEMOTRODE BRIDGE VP |
|
CHEMOTRODE, CHEMOTRODE
ORP
|
|
|
pH: 0…14
0…130 °C
Max. 6 bar
Refillable electrolyte
Pt100 in VP-version
- The problem solving pH sensor, that
just works longer!
- SKYLYTE electrolyte, air pressurizable for a poison-resistant
liquid junction
- EVEREF-B reference cartridge
for highest protection of the reference half cell and silver
free electrolyte
- 1 platinum diaphragm for clogfree
operation
- HAMILTON “PHI” pH glass
- H-XL-type comes with extra thick
“H” pH glass for abrasive samples
|
|
|
pH: 0...14
0…130 °C
Max. 6 bar
Refillable electrolyte
- Viscous 3 M KCl electrolyte “KCl-LR”
air pressurizable for less contamination
- EVEREF-F reference cartridge
prolongs electrode life in aggressive samples
- 3 High Performance ceramic
diaphragms for lower risk of blocking
- CHEMOTRODE ORP is the
Redox version and comes with a huge platinum ring
- Fits into most common
pressurizable armatures for electrodes with a reservoir
diameter of about 30 mm
|
|
|
pH: 0…14
0…130 °C
Max. 6 bar at 130 °C
Pt100 in VP-version
- It’s a POLISOLVE (pg. 9)!
- CIP, steam sterilizable
and autoclavable
- HAMILTON “PHI”
pH glass is almost impossible to poison from process samples
- Most reproducible measurements
even in hot caustic due to proprietary pH glass/low drift
- Two SINGLE POREs for direct
sample contact with the POLISOLVE electrolyte. No clogging!
- Upside-down mounting with
VP-types possible
|
|
|
| pH: Chemical Industry, Hydrofluoric Acid |
MECOTRODE, MECOTRODE
VP,
OXYTRODE PT |
|
POLILYTE HT, POLILYTE
HTVP,
POLILYTE HTSK |
|
|
pH: 0…14 or
Redox: +/- 2000 mV
0…130 °C; Max. 16 bar at
25 °C or 6 bar at 130 °C
Pt100 in VP
- Maintenance free reference
gel electrolyte
- 3 High Performance ceramic
diaphragms for reduced flowing potentials when mounted in
pipes
- Hamilton “H”
pH glass provides the most accurate readings at high pH
values or high temperatures
- OXYTRODE PT is a Redox
sensor with a platinum wire coil melted onto the glass
- Standard 12 mm shaft
|
|
|
pH: 0…14
0…130 °C
Max. 6 bar at 130 °C
Pt100 in HTVP and
HTSK version
- It’s a POLISOLVE (pg. 9)!
- Withstands continuous
high process temperature
- Specially formed pH glass
for improved mechanical stability
- Propriety pH glass for
improved mechanical stability
- Two SINGLE POREs for direct
sample contact with the POLISOLVE electrolyte. No Clogging!
- Upside-down mounting with
VP-types possible
|
|
|
pH: 0…14
-10…100 °C
Max. 6 bar at 100 °C
For up to 0.01M HF at 20 °C
and 0.05M HF at 50°C
- It’s a POLISOLVE (pg. 9)!
- HF - Hydrofluoric acid
containing applications dissolve normal pH glasses within
hours at low pH values
- HAMILTON "HF"
pH glass is much more resistant to HF than other pH glasses.
This significantly increases sensor lifetime
- Stable readings in those
harsh conditions
- Two SINGLE POREs for direct